ISO 8600-2-2015 pdf free download.Endoscopes – Medical endoscopes and endotherapy devices — Part 2:
Particular requirements for rigid bronchoscopes
Endoscopes – Endoscopes inédicaux et disposit ifs dendothéra pie — Partie 2: Exigences particulières pour bronchoscopes rigides.
4.3.1 A rigid ventilation bronchoscope shall be provided with an integral ventilation connector and an end-cap at its proximal end. Provision shall be made for insertion of endoscopes and endoscopic accessories through the end-cap. The end-cap shall not accidentally detach at an airway pressure of less than 4,0 kPa (40 cmH2O).
An end-cap may incorporate a transparent window and/or may permit the insertion of an endoscope or endoscopic accessory through the opening or an airtight gasket. Removal or opening of the end-cap for accessory insertion may be achieved by the detachment, rotation or sliding of the end-cap.
4.3.2 A rigid ventilation bronchoscope shall be provided, In a side-arm, either with an Integral ventilation connector or shall be provided with a detachable adapter which shall meet the requirements for 15 mm male conical connectors specified in ISO 5356-1 (see Figw-e 1). This shall not be a connector in which the internal lumen has been reduced. The ventilation connector, when fitted to the rigid ventilation bronchoscope, shall permit to-and-fm ventilation of the patient.
Adding a smaller connector to a 15 mm adapter creates a low dead-space connector, Users should be aware of the danger of using two connected low dead-space connectors, which may cause excessive flow resistance.
A ventilation connector may be located at either side of the rigid bronchoscope (see Figure 1) and may swivel around the shaft of the rigid bronchoscope.
4.4 Rigid jet-ventilation bronchoscopes
4.4.1 RIgid jet-ventilation bronchoscopes shall be provided with a jet-injector intended for jet ventilation.
Jet ventilation through a rigid bronchoscope is usually possible without a separate jet-injector port. Figure 2 gives examples of different bronchoscopes where jet injection is possible.
4.4.2 When tested in accordance with 5.3, a jet-injector supplied with a bronchoscope shall withstand a minimum force of 20 N without becoming detached.
4.4.3 When tested in accordance with 5j the container pressure generated by a rigid jet-ventilation bronchoscope with any jet-injector provided by the manufacturer shall not exceed 6,0 kPa (60 cmH2O).
If the bronchoscopist uses jet-injectors other than those provided by the manufacturer, he/she is responsible for its function and performance.
The ventilation pressure developed by any jet-injector depends on the characteristics of the bronchoscope with which It is used or tested. Users should be aware that ventilation pressure can be greatly increased by insertion of any obstruction, e.g. an aspirator or forceps, into the lumen of the rigid bronchoscope.ISO 8600-2-2015 pdf free download.

Download Information
Go To Download

All resources on this website are from the Internet. It is only for personal study and research. No one is allowed to use it for any commercial purpose. If you need, please buy the genuine standard. If there is any infringement of your copyright, please contact the webmaster in time, and I will correct it immediately.