IEC 61326-1-2020 pdf free download.Electrical equipment for measurement, control and laboratory use — EMC requirements — Part 1: General requirements
Materiel électrique de mesure, de commande et de laboratoire — Exigences relatives a Ia CEM — Partie 1: Exigences générales.
6.4.1 General
The general principles (performance criteria) for the evaluation of the immunity test results are the following. Permissible LOSSES OF PERFORMANCE are allowed only if specified in the test plan before testing, and clearly provided to the user in the product specification. If the PERFORMANCE LEVEL or the permissible LOSS OF PERFORMANCE is not specified, either of these may be derived from the product specification or user documentation and what the user may reasonably expect from the equipment if used as intended.
6.4.2 Performance criterion A
The equipment shall continue to operate as intended during and after the test. No DEGRADATION OF PERFORMANCE or LOSS OF FUNCTION is allowed below a PERFORMANCE LEVEL specified in the user documentation, when the equipment is used as intended. In the case of applying immunity tests with continuous electromagnetic phenomena, the PERFORMANCE LEVEL may be replaced by a permissible LOSS OF PERFORMANCE which shall recover, without user intervention. A permissible LOSS OF PERFORMANCE is allowed within the PERFORMANCE LEVEL only when this information is clearly provided to the end user via documentation, such as the product user manual. No change in the operating state is allowed nor is loss of data.
6.4.3 Performance criterion B
The equipment shall continue to operate as intended after the test. No DEGRADATION OF PERFORMANCE or LOSS OF FUNCTION is allowed below a PERFORMANCE LEVEL specified in the user documentation, when the equipment is used as intended. During the test, the equipment PERFORMANCE LEVEL may be replaced by a permissible LOSS OF PERFORMANCE if such LOSS OF PERFORMANCE is detailed in the EMC test plan. A permissible LOSS OF PERFORMANCE is allowed within the PERFORMANCE LEVEL only when this information is clearly provided to the end user via documentation, such as the product user manual. An unintended change of the operating state is allowed if self-recoverable. No loss of stored data is allowed.
The following are examples of performance criterion B:IEC 61326-1-2020 pdf free download.

Download Information
Go To Download

All resources on this website are from the Internet. It is only for personal study and research. No one is allowed to use it for any commercial purpose. If you need, please buy the genuine standard. If there is any infringement of your copyright, please contact the webmaster in time, and I will correct it immediately.