Position:Home>IEC Standards>IEC 62321-10-2020 pdf free download

IEC 62321-10-2020 pdf free download

IEC 62321-10-2020 pdf free download.Determination of certain substances in electrotechnical products Part 10: Polycyclic aromatic hydrocarbons (PAHs) in polymers and electronics by gas chromatography-mass spectrometry (GC-MS)
Determination de certaines substances dans les produits électrotechniques Partie 10: Hydrocarbures aromatiques polycycliques (HAP) dans es polymères et les produits électroniques par chromatographie en phase gazeusespectrométrie de masse (GC MS).
a) 1 000 mg/I each standard mixed stock solution:
The standards of Table 2 are respectively weighed to 0,1 g (100 mg) with an accuracy of 0,001 g, then placed in a 100 ml beaker, dissolved in a small amount of dichloromethane (Clause 5 a)), and measuring the volume of 100 ml after they are filled up with dichloromethane graduated to the mark and transferred to a flask (Clause 6 b)) in order to well shake the mix. (If necessary, it may be shaked ultrasonically.)
b) Intermediate standard mixed solution of 20 mg/I for GC-MS analysis:
Pipet 2 ml from each stock standard solution (8.4.3 a)) and 2 ml surrogate standard solution (8.4.5 b)) into a 100 ml volumetric flask and fill with dichloromethane up to the mark.
c) Intermediate standard mixed solution of 10 mg/I for GC-MS analysis:
Pipet 1 ml from each stock standard solution (8.4.3 a)) and I ml surrogate standard solution (8.4.5 b)) into a 100 ml volumetric flask and fill with dichloromethane up to the mark.
d) For low concentration samples, test standard solutions for GC-MS analysis:
For the preparation of low concentrations of the calibration standard solution (8.4.3 e) to 8.4.3 h)), PAH standard solutions are prepared for GC-MS analysis as shown in Table 3. These standard solutions each contain (20, 50, 100, 200) pg/I of one of the 18 PAH compounds. Internal standards each contain 50 pg/I of substance (such as naphthalened8, anthracene-dlO and benzo[a]pyrene-d12).
e) 20 pg/I calibration standard solution:
Pipet 0,02 ml from the 10 mg/I standard solution and 0,1 ml from the 5 mg/I working internal standard solution into a 10 ml volumetric flask and fill with dichloromethane up to the mark.
f) 50 pg/I calibration standard solution:
Pipet 0,05 ml from the 10 mg/I standard solution and 0,1 ml from the 5 mg/I working internal standard solution into a 10 ml volumetric flask and fill with dichloromethane up to the mark.
g) 100 pg/I calibration standard solution:
Pipet 0,1 ml from the 10 mg/I standard solution and 0,1 ml from the 5 mg/I working internal standard solution into a 10 ml volumetric flask and fill with dichloromethane up to the mark.IEC 62321-10-2020 pdf free download.

Download Information
Go To Download

All resources on this website are from the Internet. It is only for personal study and research. No one is allowed to use it for any commercial purpose. If you need, please buy the genuine standard. If there is any infringement of your copyright, please contact the webmaster in time, and I will correct it immediately.